Gammex Inc.
Gammex Inc.
Gammex Home

Laser Alignment Products >
Probe+ GLD450 and Probe+GLD250

   Go BackView Catalog  Print

Probe+

Diode Lasers

GAMMEX GLD 250 SAGITTAL

GAMMEX GLD 450 CROSSHAIR

 

 

The Gammex Probe+ lasers are designed with a 160 degree turret rotation.  Often, existing architectural configurations limit

sites from installing traditional lasers, but with the Gammex Probe+ difficult room designs are no problem. Our unique and

proprietary thermal design of the Probe+ series ensure users of no visible “drift”. Vertical and horizontal adjustments allow for

angular and planar movement making the Probe+ system easy to install.  The Probe+ laser alignment systems use a 635 nm

wave length laser light and built-in controls to vary the light intensity. Most of the diode products currently available for patient

alignment are in the 650 to 670 nm range. Because of the optical response of the human eye, the Probe+ line is perceived

as four times as bright as the 670 nm, and two times as bright as the 650 nm. The 635 nm yields essentially the same

brightness that you can obtain with our helium neon (HeNe) lasers. The Probe+ also allows a variable intensity control that

allows you to adjust the laser brightness based on the room lighting conditions.  Another benefit Gammex lasers offer is a

protective shielding for the electronics and especially the diode.  This enables our lasers to last longer and perform better

than other similar lasers on the market. The Probe+ series also have the same width base plate as the Exact-Align system

so it can be retrofitted to existing turret laser installations with no modifications.  This makes conversion from the HeNe

systems to the Diode systems cost friendly and easy.  The Probe+ diode laser has a 0.5 mm line at a distance of 2 meters.

Complete documentation on laser beam profile is included with each Probe+ diode alignment system.  The standard Probe+

Diode laser utilizes an external power supply adaptable for international use.  The Probe+ Diode laser system is easy to

repair and maintain.

 

SPECIFICATIONS

Laser Beam Output

Power . . . . . . . . 0.5 mW (each beam)

Range . . . . . . . . Up to 6.0 m

Line Width . . . . . <0.5 mm @ 2 m

Line Divergence (Length). . . . . . . 15-17°

Drift . . . . . . . . . No measurable drift

WaveLength . . . . 6350 Å (635 nm) visible red

Visibility . . . . . . Clearly visible in strong ambient light

Laser Beam Adjustment

Horizontal Range of Vertical Projection . . . . . 330.2 cm @ 3.05 m

Vertical Range of Horizontal Projection . . . . . Plus 100 cm /minus 300 cm @ 3 m

Line angle adjustment range . . . . . . . . ±180 degrees

Mounting . . . . . . Turret adjustable to ±80°

Laser Dimensions

Length. . . . . . . . 17.3 cm (6.75 in)

Width . . . . . . . . 14.0 cm (5.5 in)

Depth . . . . . . . . 8.0 cm (3 in)

Weight. . . . . . . . 1.8 kg (4 lbs)

Power

Requirements . . . 12 VDC, 300 mA

Universal Power Supply

Power Requirements . . . 110 VAC – 220 VAC, 50 – 60 Hz, 0.4 A

Laser unit includes various adapters for U.S. and international use.

Certification

Complies with Center for Devices and Radiological Health regulations for Class II Lasers.

Warranty