Mammographic
DCF Test Tool
GAMMEX 150K |

|
|
The Gammex 150K Mammographic Density Control Function (DCF) test tool enables a quick and accurate assessment of
a film-screen mammography unit’s Automatic Exposure Control (AEC) accuracy.
The Mammographic DCF Test tool can record up to eleven exposures on a single piece of film. This saves film cost, set-up
time, and processing time. Combining all eleven exposures on a single piece of film eliminates processing variability. The
resulting density read-out can be used to perform ACR Density Control Function Tests.
The Gammex 150K consists of a base plate with an engraved density scale and a sliding exposure plate with a small
window. By sliding the exposure plate without moving the cassette, the same film can be exposed at each of up to eleven
AEC density steps (-5 to +5). Lightweight, convenient, and economical, the Gammex 150K Mammographic DCF Test Tool
is an excellent addition to your mammographic QA program. |
|
|
SPECIFICATIONS |
|
Construction
Base Plate . . . . . Aluminum
Exposure Plate . . Stainless Steel
Dimensions
Overall . . . . . . . 15.2x30.5x.95 cm (HWD) (6x12x.375 in)
Exposure Window . . . . . . . 12x18.6 mm (.47x1.12 in)
Weight. . . . . . . . 0.8 kg (1.75 lbs)
Film size . . . . . . 7x9.5 in
Features
Exposure Steps . . 11 (-1 to -5, zero, +1 to +5)
Compatibility . . . Tissue Equivalent Breast material and acrylic (Gammex 159, 159A,
159BR and 159A-BR Phantoms)
NOTE: Tissue Equivalent Material sold separately | |
|
Optional Carrying Case model 150K-CASE is available. |