Gammex Inc.
Gammex Inc.
Gammex Home

Mammography Products >
Gammex 157A Mammographic Film/Screen Contact Test Tool

   Go BackView Catalog  Print

Mammographic Film/Screen

Contact Test Tool

GAMMEX 157A

 

Film/Screen contact can have a significant influence on image quality.  Today’s mammography film/screen systems have

higher resolution than diagnostic radiography x-ray systems and require test tools with finer detail.  Regular testing with the

Gammex 157A detects problems and artifacts early. Use of the tool and the evaluation of the resulting images is simple.

Areas of poor screen contact appear darker than areas of good contact. Any dark areas greater than 1 cm should be

investigated and corrective action taken. Gammex recommends testing cassettes every six months or when new or repaired

cassettes are put into service.

 

 

SPECIFICATIONS

 

Screen Size. . . . . 24x30 cm (9.4x11.8 in)

Mesh . . . . . . . . . #40 Mesh - Copper

Size. . . . . . . . . . 25.8x31.5 cm (10.2x12.4 in)

Weight. . . . . . . . 0.4 kg (0.9 lbs)