| Digital kV, Dose and Time Meter, Gammex 330 |
Digital kV, Dose and Time Meter
Gammex 330
|
Product Overview
The Gammex 330 Digital kV, Dose and Time Meter is a test device for quality control and acceptance testing in radiographic, mammographic and fluoroscopic x-ray systems.
|
Benefits
- Measures kVp, dose and time non-invasively
- Can be used for both acceptance testing and routine quality control
- Interface permits printout of results or online storage
|
|
Features
- Includes digital display of the quantity PPV (pratical peak voltage) according to IEC 61676
- Compact and light-weight
- Easy-to-read LC Display
- Measures kVp, dose and time non-invasively
- Touch key controls
|
Certification and Warranty
- The Digital kV meter is manufactured for Gammex and carries the standard warranty of its manufacturer, one year.
|
|
SPECIFICATIONS |
|
Tube Voltage Selectable calculations . . . . PPV, max kVp, mean kVp Measuring range . . . . . . . . . 40 - 150 kV, 22 - 40 kV (MAM) Dose . . . . . . . . . . . . . . . . . . . 50 µGy - 50 Gy; 150 µGy - 150 Gy (MAM) Time . . . . . . . . . . . . . . . . . . . 0.3 ms - 999 s Digital Resolution . . . . . . . . 0.1 kV Accuracy . . . . . . . . . . . . . . . . ±1% or ±1.0 kV; ±0.7kV (MAM) Interface . . . . . . . . . . . . . . . . RS232 and analog kV waveform Power Supply . . . . . . . . . . . . 4 NiMH batteries (AA) Dimensions . . . . . . . . . . . . . . 45x95x155 mm (HWD) (1.75x3.75x6 in) Weight . . . . . . . . . . . . . . . . . . approx. 770 g (1.7 lbs) without batteries
|
|
|

|
|